Brief Introduction:
YS6002 benchtop color haze meter is developed by 3nh independently with proprietary intellectua property. This machine adopts transmittance optical geometry d/0 (diffused illumination, 0-degree viewing
angle) which confirms to CIE 15 standard. It can also realized the the optical geometry 0/0 (parallel illumination, 0-degree viewing angle).
YS6002 color haze meter uses 360~780nm composite full spectrum LED light source (which is closed to the spectrum of D65), and it can realize intervals 10nm spectrum by precision concave grating and 256 Image Element CMOS Sensor. Accordance with industry-standard, YS6002 will give the accuracy result for the measurement of haze, transmittance, chromaticity, color difference, WI, YI, Pt-Co Index and Gardner index etc. It is wildly used in the field of glass processing, plastic processing, film manufacturing, display screen processing and packaging industry, liquid pharmaceutical analysis, laboratory etc.
Features:
a. High configuration for the handware: 7 inch capacitive touch screen; 25000 data storage ; with USD / Bluetooth data port.
b. With high optical configuration and resolution, 154mm integrating sphere, concave grating, 256 Image Element CMOS Sensor, dual light path;
c. Adopt 360~780nm composite full spectrum LED light source (which is closed to the spectrum of D65), which is low power consumption and long life time.
d. Easy realize the precision measurement for 3 kinds of light sources A,C and D65 for haze (ASTM D1003/1044), transmittance, Pt-Co Index and Gardner index.
e. Multiple measurement modes : quality control mode and sample mode to meet more customers’ requirement;
f. Independent light source detector and built-in temperature sensor will continuously monitor the change of light sources and environment to ensure the measurement more reliable.
g. Observer angle 2/10, multiple light sources, multiple color spaces, multiple chromaticity to meet different kinds of customers’ testing requirements.
h. Good man-machine interactive to ensure the measurement accuracy and stable;
i. More powerful extended functions at the PC software.
Model | YS6002 |
Optical Geometry |
Transmittance:d/0 (diffused illumination, 0-degree viewing angle); similar to CIE15 standard 0/0(parallel illumination, 0-degree viewing angle); Confirms to the standard of ASTM D1003/1044, CIE No.15, GB/T 3978, GB 2893, GB/T 18833, ASTM E308, DIN5033 Teil7 |
Application |
wildly used in the field of glass processing, plastic processing, film manufacturing, display screen processing and packaging industry, liquid pharmaceutical analysis, laboratory etc. |
Intergrating Sphere Size | Φ154mm |
Light Source | 360~780nm Combined LED Lamp |
Spectral Separation Device |
Concave-Grating |
Detector | 256 Image Element Double Array CMOS Image Sensor |
Wavelength Range | 360~780nm |
Wavelength Pitch | 10nm |
Half Bandwidth | 10nm |
Reflectance Range | 0~200% |
Measurement Aperture | Transmittance: Φ25.4mm |
Color Spaces | CIE LAB,XYZ,Yxy,LCh,CIE LUV, Musell,s-RGB, HunterLab,βxy,DIN Lab99 |
Color Difference Formulas |
ΔE*ab,ΔE*uv,ΔE*94,ΔE*cmc(2:1), ΔE*cmc(1:1),ΔE*00, DINΔE99,ΔE(Hunter) |
Other Colorimetric Data |
Haze (ASTM D1003/1044),transmittance rate T, WI(ASTM E313,CIE/ISO,AATCC,Hunter), YI(ASTM D1925,ASTM 313), TI(ASTM E313,CIE/ISO), Mt(Matamerism Index ), Color Strength, Staining fastness,Color Strength, Opacity, Pt-Co index and Gardner index |
Observer Angle | 2°/10° |
Illuminant | D65,A,C,D50,D55,D75, F1,F2,F3,F4,F5,F6,F7, F8,F9,F10,F11,F12 |
Display Data |
Spectral/Values, Chromaticity Values, Color Difference Values/Graph,Color Diagram, Color Simulation, Pass/ Fall Result, Color Offset |
Measurement Time | About 1.5s |
Repeatability |
Spectral transmittance: for test transmission reference white within ΔE*ab 0.05% (400~700nm: within 0.04%); Chromaticity Data: Φ25.4mm,within ΔE*ab 0.03 (after do warm-up and calibration, it is the average value to measure transmission reference white plate for 30 times at 5 second intervals) |
Inter-instrument Agreement |
Φ25.4mm,within ΔE*ab 0.15 (average value for 5 pcs standard transmittance color plates measurement ) |
Size | 370X300X200mm(L*W*H) |
Weight | About 9.6kg |
Power | DC 24V,3A Power Adapter |
Illuminant Life Span | 5 years, more than 3,000,000 times measurement |
Screen | 7-inch TFT Capacity Touch Screen Display |
Interface | USB,Printing Port |
Storage | Standard:5000; Sample:20000 |
Language | Simplified Chinese, Traditional Chinese, English |
Working Temperature | 0~40℃(32~104°F) |
Storage Temperature | -20~50℃(-4~122°F) |
Standard Accessory |
Power Adapter, User Manual, QC software(please download from website), USB cable, reflectance white and black board, transmittance black board, 25.4mm aperture, transmissive test component |
Optional Accessory | Miniature- printer |
Note | The specifications are subject to change without notice. |
With 12 rigorous quality inspection processes in place, we ensure a high yield rate of up to 98% for batch processed precision parts.
FAQs About Haze Meters
D/8°and 45°/0°are two kinds of illumination geometry with different optical geometry design. D/8°illumination geometry is more widely used for the color comparison and measurement of high-gloss materials; 45°/0°illumination geometry is mainly used in printing and packaging industries.
There are three measuring apertures for NH310: Φ8mm (standard accessory), Φ4mm (standard accessory), extended Φ8mm (optional accessory)
a. Turning on
b. Removing measuring aperture, install the needed measuring aperture.
c. Select "Settings-Aperture Setting" in main menu to select corresponding measuring aperture.
d. After selecting aperture, the instrument will display "White and Black Calibration" interface. White and black calibration must be performed.
e. Replace completed.
Since part of the light will pass through the light-transmitting materials, the colorimeter is not suitable for color measurement of such kinds of materials; however, when place a pure color material with uniform texture under the light-transmitting material, we can measure the color difference between the standard and the sample.
Threenh Technology colorimeter is shipped with an operating instructions and video of English version.If you need an electronic version of the user manual, you can download it from our website download center
a. Check whether the USB cable is connected. If not, please connect it. Check whether the connection is effective. You can re-plug the USB cable to test.
b. If the USB cable is connected well but the interface still displays "USB Disconnect", please re-start the colorimeter and enter "Menu" -"Comm"to start the communication.
c. Change the USB cable or change the USB port in you PC.