11-Steps Gray Scale Test Chart

This 11-Steps Gray Scale Test Chart Instrument are Certified with
certify
1 Years Warranty (Additional support period of 3 years)

The YE0165 is designed for evaluation the halftone reproduction of 16:9 electronic cameras. Two 11-graduated counter current gray scales are arranged on a gray background (D ≈ 0.56), the gray scale corresponds to the ITU recommendation BT.709.

The output signal of an optimally gamma-corrected camera according to BT.709 generates two counter current, linear step signals.

The contrast range of the gray scale is 40 : 1.

The values of the 11-graduated gray scale are as follows:

Step Density Remission in % Video signal for BT.709 in % Video signal for gamma 0.45 in %
1 0.05 89 100 100
2 0.13 74 91 92
3 0.22 60 82 84
4 0.32 48 73 75
5 0.43 37 64 67
6 0.56 28 55 59
7 0.70 20 46 51
8 0.88 13 37 43
9 1.08 8 28 35
10 1.34 5 19 26
11 1.64* 2 11 19

*The production process for reflective charts allow a maximum printed density of 1.64. The target value for the darkest step is 1.70. Therefore the distance of the last step differ slightly.

Two black fields and a white field are located between the gray scales. The density of the black fields in the middle is D > 3 (remission < 0.1 %), the white field between has a density of 0.05 (remission 89.9%).
Recommended usage:

Adjust the aperture until the lightest patch reaches 100% video signal at the waveform monitor. With an optimally corrected camera according to ITU recommendation BT.709 the darkest patch will reach 10% signal level.

When using this chart with a gamma correction of 0.45 the video signal level for each step will change as listed in the table above.

Our Valued Clients

FAQs About  Grayscale Charts

Who are we? Threenh Technology: 22-Year Professional Color Measurement Solution Provider

Threenh Technology is a Chinese professional Measurement Instrument manufacturer with 22 years of industry experience, specializing in the field of color measurement and surface gloss detection. Integrating research and development, design, production, marketing, and sales, the company takes technological innovation as its core and is committed to providing global customers with high-precision and intelligent testing instruments as well as professional solutions.


Gloss level is not given out in percentage but in gloss units (GU). In practice, however, 100 GU is considered 100 percent reflective. To contrast visually, the 20-40 GU is a low-gloss surface, and 85 or more is almost 100 percent mirror-like reflection.


The repeatability is an important parameter for the accuracy of the colorimeter. Threenh Technology portable colorimeter for color measurements tandard deviation ΔE*ab is within 0.03 (Measurement condition: the average value of 30 measurements on the white board at an interval of 3s after calibration). Each colorimeter can pass the qualified inspection before leaving the factory to ensure it conforms to the standard of China Metrology Institute.

The main instruments used to detect color are spectrophotometers and colorimeters (including photoelectric integrating colorimeters). 

Spectrophotometer: High-precision option. It analyzes the full visible light spectrum to measure color accurately. Suitable for complex scenarios like textured surfaces, special effect colors, or batch consistency checks in industries such as paint and coatings. 

Colorimeter (Photoelectric Integrating Colorimeter): Cost-effective and portable. It uses RGB filters to measure tristimulus values directly. Ideal for quick color difference detection in simple applications. Key Selection Tip Choose based on accuracy needs: use a spectrophotometer for high-precision color measurement, and a colorimeter for fast, basic color difference checks.


D/8°and 45°/0°are two kinds of illumination geometry with different optical geometry design. D/8°illumination geometry is more widely used for the color comparison and measurement of high-gloss materials; 45°/0°illumination geometry is mainly used in printing and packaging industries. 

SCI&SCE are two method in the color measurement. SCI means Specular Component Include,SCE means Specular Component Exclude.

Under the method of SCE, only test diffuse refection and exclude specular reflection. In that way, the test result is similar to object color was observed by human eyes.

Under the method SCI, both the diffuse refection and specular reflection will be included. In that way, the value about the color is more objective. It will not effect by the environment condition.

When we choose the instrument, those elements should be taking into consideration.


Related Grayscale Charts You May Also Like

Give us a call!

Discuss your project with one of our product specialists:

0086 755 26508999

Need pricing?

We respond to quote requests within 1 business day.

service@3nh.com

Already know what you need?

Submit a demo request for the product you're interested in.

Get a Demo
Leave Your Message
We use cookies to ensure that we give you the best experience on our website. If you continue to use this site we will assume that you are happy with it. Privacy policy